Appendix IV : Labpro instrumentation spec.s

These units are Data Acquisition Devices (DAQ) they record voltages (range =±10V, resolution = 4.9mV) from attached sensors at programmed intervals or other trigger conditions.  They may also convert the raw data to measurements, average it, or filter it.

 

 

LabPro

On-board Data Display

No

Max # of data Points

<12,287 (limited by memory)

Sample Time Range (seconds/ sample)

10-4s to 16,000s

Sample Time Rules

Multiple of 100ms

Max # of probes: Analog vs Digital

4 Analog, 2 digital

Data collection and analysis: Realtime vs Non-realtime

both

Setup: Remote, Calculator, Computer

All three modes

Connection: I/O port (Calculator to calculator link cable), Serial port, USB port

All three

 

 

 

Software

DataMate : A calculator program (transferred from LabPro) that allows any probe or combination of probes to be setup, calibrated, collected and graphed.

Logger Pro: A computer program that will setup, calibrate, collect, graph, store, and analyze data from all probes on the LabPro.

TI-Graph Link: A computer program that allows programs and data to be transferred between the calculator and the computer.  It requires a special adapter cable from the computers serial port and the calculator's I/O port.

Graphical Analysis: A computer program that allows data to be retrieved directly from the calculator and then graphed and analyzed.

Sensors

Many sensors are available.  They return a voltage (V) that is linearly proportional to the desired measurement (M).  This voltage must be "calibrated" by getting a slope (K1) and intercept (K0) for the following equation. 

 

 

 

 

 

 

 

Standard Calibration Coefficients:

Sensor

ID

Units

Equ Type

K0

K1

K2

Stainless Steel Temp

TMP-BTA

°C

12

0.001021

0.0002225

0.0000001333

Barometer

BAR-BTA

Atm

1

0.8093

0.077

 

 

 

mBar

1

819.52

78.001

 

 

 

mmHg

1

614.84

58.52

 

 

 

inHg

1

24.215

2.292

 

CO2 Sensor

CO2-BTA

ppm

1

0.0

2000

 

O2 Sensor

O2-BTA

%

1

0.0

6.658

 

Relative Humidity

RH-BTA

%

1

-23.8

32.9

 

Magnetic Field (low)

MG-BTA

Gauss

1

-80.63

32.25

 

 

 

mTesla

1

-8.063

3.225

 

Light Sensor      –600 lux

LS-BTA

lux

1

0.0

154.0

 

                        -6000 lux

 

lux

1

0.0

1695.0

 

                        -150000

 

lux

1

0.0

38424.0

 

Radiation Monitor (G-M)

RM-BTA

Counts

1

0.0

1.0

 

 

 

 

 

 

 

 

 

Sampling Interval Limitations:

LabPro           S cannot be less than 0.0001 sec nor more than 16,000 seconds.  S must be a multiple of 0.0001 sec. 

TI-83 specs:

Memory:         27K bytes

Lists:               L1 thru L6, plus unlimited additional named lists (5 chars).  Each list may contain up to 999 elements.  However, since 9bytes are required per element, one is usually limited to about 512 data points because of memory.