SN 1109 Pre-Encap summary


SN
1109
Status
Pre-Encap
Test Result
Fail
Operator
Mutsumi
TE_Config
009
Software_V
rel-1-3-4
Reference MCM S/N
1077
Location
Teledyne
GTRC version
7
Board type
Tall(899)
Start Time
2004/05/11 13:22:33
End Time
2004/05/11 13:31:32
Dead or Hot channels
[]
Bad Gain or Noise channels
[]

Power Consumption and LVDS levels

Functionality

Bias leakage current (TM703)

Bias_V (V) Bias_I (nA)
2.0 0.6134
4.0 1.0644
6.0 1.6367
8.0 2.0840
10.0 2.6913
15.0 5.7687
20.0 7.8665
Bias_V (V) Bias_I (nA)
30.0 20.5178
40.0 24.6665
50.0 33.2850
60.0 36.2432
70.0 36.9792
80.0 38.2836
90.0 40.8786
Bias_V (V) Bias_I (nA)
100.0 42.1120
110.0 42.5250
120.0 41.2175
130.0 44.1304
140.0 44.2547
150.0 44.8618
160.0 44.1262
Bias_V (V) Bias_I (nA)
170.0 47.0840
180.0 48.1562
190.0 46.5129
200.0 48.1960

Threshold scan without charge injection (TM710)

Gain and Noise (TM714)

Noise trigger rate threshold scan (TM715)