SN 1099 Pre-Encap summary


SN
1099
Status
Pre-Encap
Test Result
Pass
Operator
rlw
TE_Config
009
Software_V
rel-1-3-3
Reference MCM S/N
1077
Location
Teledyne
GTRC version
7
Board type
Short(898)
Start Time
2004/05/04 11:16:12
End Time
2004/05/04 11:33:57
Dead or Hot channels
[]
Bad Gain or Noise channels
[]

Power Consumption and LVDS levels

Functionality

Bias leakage current (TM703)

Bias_V (V) Bias_I (nA)
2.0 0.6477
4.0 1.2363
6.0 1.9415
8.0 2.5622
10.0 3.3543
15.0 6.8214
20.0 9.0159
Bias_V (V) Bias_I (nA)
30.0 22.1360
40.0 30.7566
50.0 36.6061
60.0 43.2137
70.0 45.7264
80.0 48.6911
90.0 51.2329
Bias_V (V) Bias_I (nA)
100.0 53.0681
110.0 54.4950
120.0 56.6735
130.0 54.6787
140.0 55.6633
150.0 56.8693
160.0 55.0822
Bias_V (V) Bias_I (nA)
170.0 55.3873
180.0 57.1290
190.0 54.8320
200.0 54.0981

Threshold scan without charge injection (TM710)

Gain and Noise (TM714)

Noise trigger rate threshold scan (TM715)