SN 235 Pre-Encap summary


SN
235
Status
Pre-Encap
Test Result
Pass
Operator
rlw
TE_Config
009
Software_V
rel-1-3-3
Reference MCM S/N
1077
Location
Teledyne
GTRC version
7
Board type
Tall(899)
Start Time
2004/05/04 14:04:10
End Time
2004/05/04 14:21:49
Dead or Hot channels
[]
Bad Gain or Noise channels
[]
Bad trigger channels
No Data
Number of bad response channels
0
Disconnected pitch-adapter channels
No Data
Number of disconnected pitch-adapter channels
No Data
Number of bad channels in total
0

Power Consumption and LVDS levels

Functionality

Bias leakage current (TM703)

Bias_V (V) Bias_I (nA)
2.0 0.6238
4.0 0.9952
6.0 1.5611
8.0 2.1525
10.0 2.7660
15.0 5.9586
20.0 8.2571
Bias_V (V) Bias_I (nA)
30.0 22.4980
40.0 30.4305
50.0 34.8517
60.0 41.0897
70.0 47.1453
80.0 49.3545
90.0 54.1561
Bias_V (V) Bias_I (nA)
100.0 55.6799
110.0 57.8766
120.0 57.8043
130.0 58.1586
140.0 59.0511
150.0 61.5574
160.0 60.4303
Bias_V (V) Bias_I (nA)
170.0 59.9040
180.0 60.2370
190.0 60.3963
200.0 59.9535

Threshold scan without charge injection (TM710)

Gain and Noise (TM714)

Noise trigger rate threshold scan (TM715)

TREQ response (TM713)

Pitch adapter signal channel connection

Pitch adapter bias connection